Towards distortion-free imaging of the eye
DOI 10.1371/journal.pone.0252876 , Volume: 16 , Issue: 6 , Pages: 0-0
Article Type: research-article, Article History
Abstract

The high power of the eye and optical components used to image it result in “static” distortion, remaining constant across acquired retinal images. In addition, raster-based systems sample points or lines of the image over time, suffering from “dynamic” distortion due to the constant motion of the eye. We recently described an algorithm which corrects for the latter problem but is entirely blind to the former. Here, we describe a new procedure termed “DIOS” (Dewarp Image by Oblique Shift) to remove static distortion of arbitrary type. Much like the dynamic correction method, it relies on locating the same tissue in multiple frames acquired as the eye moves through different gaze positions. Here, the resultant maps of pixel displacement are used to form a sparse system of simultaneous linear equations whose solution gives the common warp seen by all frames. We show that the method successfully handles torsional movement of the eye. We also show that the output of the previously described dynamic correction procedure may be used as input for this new procedure, recovering an image of the tissue that is, in principle, a faithful replica free of any type of distortion. The method could be extended beyond ocular imaging, to any kind of imaging system in which the image can move or be made to move across the detector.

Introduction

Raster-based scanning systems such as scanning light ophthalmoscopy (SLO) and optical coherence tomography (OCT) have become ubiquitous in both clinical practice [13] and ophthalmic research [46]. Acquired images of ocular tissue are built up over time by raster scanning across the field. Unfortunately, because the eye is in rapid and constant motion [7], no two images look quite the same. This can be described as “dynamic” distortion, which would not occur were the eye stationary. Great care is taken in research applications especially to remove or minimise dynamic distortion where possible [8,9].

We recently described a method to remove, in principle, all dynamic distortion for the scanning light ophthalmoscope [10]. The method was first suggested some time ago [11] and has been confirmed to work with OCT data in three dimensions [12,13]. The method makes a key assumption that the fixational motion of the eye is random in nature, so that if a sufficient number of frames are observed there should be no expected bias in the distortion seen for any particular piece of tissue compared with the rest of the tissue. Where this assumption holds true, this procedure can be expected to remove dynamic distortion entirely. The method has been adopted by several imaging groups [1217].

Despite its advantages, the dynamic correction method is entirely blind to distortion which manifests in the same way on each acquired frame. We refer to this as static distortion. Such distortion will occur in any ophthalmic imaging system due to marked anisoplanatism which results from the high power of the eye and of the optical systems used to image it, which is one reason that adaptive optics systems are typically limited to 1–2° in diameter [18,19]. Major distortion also arises when imaging over an extended field, due to the typically flat shape of detectors compared with the highly curved retina. Static distortions can also occur for raster systems employing resonant scanners due to error in the desinusoiding process [20]; from systematic errors of the hardware [21]; or from eye movement patterns such as nystagmus which are large, repeatable and potentially have periodicity similar to the imaging frame rate such that pseudo-static distortion could result.

To our knowledge the problem of correcting for static distortion does not appear to have been canvased in the ophthalmic imaging literature. It is difficult to state precisely how significant a problem this is, due to a lack of experimental data available on distortion of the eye [22], let alone the combination of eye and imaging system. For the eye alone, schematic eye models can be used to provide some estimate. Over a typical field of view of ~30° in a conventional ophthalmoscope pointed at the optic nerve head (15° off-axis), distortion from one edge of the field to the other is predicted at ~6% [22]. This is 1.8°, approximately one third of the diameter of the optic nerve head. Distortion is likely to be even higher a) when including the optics of the imaging system, which lack the excellent wide-field imaging properties afforded by the gradient index of the crystalline lens [19]; b) with an adaptive optics correction in place, as off-axis aberrations typically accrue more rapidly across the field than without adaptive optics [23]; and c) when constructing montages of the retina or imaging with very wide field devices [24], which can span >180°.

As opposed to the field of ophthalmic imaging, there has been extensive work on the problem of static distortion in the field of “camera calibration” in computer vision research, where it forms part of the broader problem of determining various fixed parameters of an imaging system in a 3D world. Parameters are typically inferred by registration of features within images acquired as the camera pans across a scene, typically by rotating it [25], which has obvious parallels with the constant motion of the eye with respect to a fixed ophthalmic imaging system.

The traditional approach to camera calibration [25] has become ubiquitious in computer vision applications [26]. Although its primary goal is to recover global features such as the camera focal length and viewpoint in the 3D world, it could in principle be applied to ophthalmic imaging. However the following limitations make this approach infeasible:

    • Most commonly this approach employs a known calibration target, but this is not possible to implement in the eye.
    • The method begins by treating the imaging system as a “pinhole” camera that is described by simple geometric transformations. By definition such a camera has no distortion. Distortion is discovered and corrected by using a parametric model for the distorting function, for example a rotationally-symmetric polynomial series. The eye, however, is not rotationally symmetric and, even if it were, off-axis viewing induces a large amount of aberrations which are not [27]; it is also necessary to solve for non-optical sources of distortion as described above, which are difficult to parameterize. A model-based approach is therefore undesirable.
    • The method typically requires that images be acquired from different rotations of the camera, with translations providing redundant information [25]. Whilst the eye certainly can exhibit torsional movement that would rotate the scene in the necessary way as it attempts to maintain steady fixation, the only motion that is guaranteed in any image stack is translation of the acquired frames.
    • To our knowledge none of the methods advanced have treated the case of dynamic and static distortions together; only static distortions have been considered.

It should be noted that various developments have been made over the traditional approach cited above, largely solving the first two problems. For example, it is possible to use the natural scene instead of a structured calibration target if the registration can be performed faithfully, i.e. common features can be located across multiple scenes. This is referred to as “self calibration”. It is further possible to solve for a complete (pixel-resolved) distortion mapping function rather than to presume some model, which may be termed a “non-parametric” calibration [28]. Various authors have made similar proposals [2932], which address the first limitation addressed above and sometimes the second, but not the other two limitations.

A form of non-parametric self-calibration has also been proposed in the field of vision science, not for imaging per se but for a visual system to compensate for distortion imposed by the optics of the eye and by irregular distribution of photoreceptor cells tiling the retina [33]. In that work it was recognised that movements of the eye shift the retinal scene and that, if the objects in the scene are actually stationary and the system well calibrated, all objects in the scene should move by the same amount. If the calibration is in error, this will no longer hold true. Therefore principled trial-and-error may be used to modify the system’s calibration function until the distortion-on-movement is minimised. This is a philosophically similar approach to the one described below, despite the different application. However, major limitations include the difficulty of ensuring convergence of the solution on a global minimum, the potentially large number of image samples required, and a “lowpass” assumption of image formation; these limitations are avoided the present work.

Here, we describe a method that is able to solve for static distortions of arbitrary form by acquisition of images from the translating (and possibly rotating) scene expected for the ever-moving eye. We show further how the previously described dynamic correction procedure can be applied to “clusters” of frames, leaving behind residual static distortion for each cluster. These partially recovered, distorted images can then be corrected by the new procedure such that, in principle, no distortion remains.

Methods

Broadly, the new procedure requires imaging the same retinal scene across a handful of frames with the eye in different positions (that is, with the images translated and/or rotated with respect to one another). One warped image is selected as a reference, against which the other images are diffeomorphically registered, such that any one of the (warped) images can be transformed to resemble the warped reference. As with the dynamic correction method [10], the approach detailed here takes as input the pixel displacements determined by successful registration; the registration itself is assumed to have been carried out successfully with some existing approach [34]. Here the registration information is used to form a system of simultaneous linear equations, with a variable for each pixel and each mapping between a pair of pixels providing an observation. Solution of this sparse system of linear equations gives the common warp field that was encountered by each frame, as detailed further below. This approach implicitly assumes that the distortion conferred by the eye itself does not change to any appreciable degree over the range of eye positions considered. This assumption is justified given that the approach requires displacements of only a few pixels (each one corresponding to ~ 0.5 μm in our images), compared with the extent of the isoplanatic zone of the eye which is typically in excess of 300 μm [18].

A note on terminology: distortion maps are defined here as a lookup table of displacements which specify how to build the output image from an input image. For example, if the distortion map in the x direction had a value of -2 at column 100, we would source the data for the pixel at column 100 from column 98 in the input image.

Matlab code and example data are included in S1 File, to assist in implementing the new algorithm.

Simplified 1D example

Consider the 1D scene shown in Table 1 containing letters of the alphabet “E” through “K”, spaced one unit apart. The scene was translated by an amount “t” pixels in the x dimension by either t = 0 units (no shift) at the time of acquisition for image 1, or t = -3 units (a shift to the left) for image 2. The translated scenes are sampled by a 1D detector that is just 3 units long. Anchored to the detector and therefore common to all images is a “static” distortion function f(x), which maps each pixel on the detector to pixel(s) in the translated image scene. In this example the distorting function was chosen to be a parabola, f(x) = x2 / 4.

table-wrap
Table 1
Simplified example in 1D.
Position (pixels)
-3 -2 -1 0 +1 +2 +3 +4 +5 +6
Undistorted space (t = 0) E F G H I J K
Static warp f(x) = x2/4 0 +0.25 +1
Acquired image #1 E FG H
Undistorted space (t = -3) E F G H I J K
Static warp f(x) = x2/4 0 +0.25 1
Acquired image #2 H IJ K
Registration map (image 2 to 1) N/A N/A -2
Registered image #2 N/A N/A H
Example consisting of a 3-unit detector sampling a scene 7 units wide. The scene is imaged twice, after translation of the scene by either t = 0 or t = -3 units. In each case a distortion map f(x) anchored to the detector re-directs the information that would have been intercepted by each pixel. The images resulting from the differing translations together with the common distortion map are shown. The final row provides a look-up table to register image 2 to image 1; it has a single entry, corresponding to the only overlapping feature between the images.

Suppose that we are tasked with determining the value of the distortion map at the pixel x = +2, using only the two distorted images together with knowledge of the overall translation of the scene. The relevant feature to this exercise, as we shall see, is the letter “H” (bolded in the Table). Other features (for example “FG”) are explained further below.

First, consider that the pixel at x = +2 on the detector “sees” a distortion value of f(2) = 22 / 4 = +1. This means that information at this pixel is sourced from one unit to the right in object space i.e. at x = +3. At the time of capturing the first image, the feature at that location is H (top row of table), and so H is recorded at x = +2 in the acquired image (third row).

Similarly, the pixel at x = 0 on the detector sees a distortion value of f(0) = 02 / 4 = 0. Therefore data is always sourced from the present pixel (x = 0) in the translated scene. At the time of capturing the second image, the feature at x = 0 is again H, and so H is recorded at x = 0 in the image.

Returning to the task of determining the distortion map, the co-ordinates in undistorted space (xU) of a given feature can be expressed independently in the co-ordinate system of the reference image (xR) and in the co-ordinate system of the image to be registered (the moved image, xM):

      In this particular example (i.e. for the feature H), these expressions evaluate as follows:

          Given the equality of these expressions (i.e. that they both refer to the undistorted co-ordinate space) we can write:

          alternatives f x R f x M = x M x R t M t R = d M , R t M , R
          where dM,R is a vector indicating the output of pixelwise registration between the two images, being equal to the difference in feature location between the two images (xM − xR). The term tM,R is a scalar representing the full-field translation of the moved image relative to the reference image.

          In other words, by mapping a feature in the reference image space (xR) to the same feature in the moved image space (xM), and adjusting for a known amount of whole field translation between the images (t), we obtain the difference in the common distorting function [f(xR) − f(xM)] between two pixels on the detector. We refer to this as a “paired functional difference ”. The terms on the right-hand side of Eq 1 are measured by registration of the two images (note: whole image translation cannot be reliably inferred from this example with only 3 pixels). The terms on the left-hand side define the (difference in the) distortion map that we are attempting to solve. By obtaining many such paired differences we can constrain the shape of the function (but not its absolute value), as illustrated in Fig 1.

          Illustration of a paired functional difference.
          Fig 1
          An example static distortion map f(x) is defined over the pixels of the detector (x). Circles represent examples where detector pixels (x0 in the moved image, x2 in the reference image) were determined, by image registration, to have “seen” the same object in different frames as was depicted in Table 1. The value of the distorting function at the two locations is f(x0), f(x2). The object in question has shifted between moved and reference frames by an amount (dM,R), which is equivalent to the sum of a) the difference in the distorting function f(x2) − f(x0), and b) the overall translation of the scene tM,R. Therefore, pixel-wise registration between shifted versions of the same scene can be used to build up information about the shape of the distorting function that is common to each frame.Illustration of a paired functional difference.

          Returning to the example of Table 1, if we acquired other images with varying translations we would possess a set of observations of paired functional differences. For example, with appropriate translations we might discover that:

          alternatives f 2 - f 0 = + 1
          alternatives f 2 - f 1 = + 0.75
          alternatives f 1 - f 0 = + 0.25

          This constitutes a system of simultaneous linear equations. In this example, the number of unknowns is equal to the number of observations so that there is no unique solution. However, we are interested only in the relative shape of f(x) since adding a constant will merely shift the entire image without distortion. This allows us to set, for example, f(x) = 0 which means here that f(1) = 0.25 and f(2) = 0.75. These match the true values of the function as shown in Table 1.

          Extension to large numbers of pixels

          Real images are comprised of many more pixels and hence there will be many paired functional differences to consider. This large set of simultaneous equations can be solved using matrix algebra as shown in Fig 2.

          Linear algebra representation required to solve for the static distortion function f(x).
          Fig 2
          A sparse array (A) populated by entries of ± 1 is used to perform pairwise difference operations between image pixels xi for the distorting function f(x). The resulting “paired functional differences” are given in Δ. This array is equivalent to the vector required to register the two images (dM,R ), after accounting for the scalar whole-field translation between the images t. The unknown values f(x) can then be solved by pseudoinversion of the large, sparse matrix A. The highlighted row corresponds to the example depicted in Table 1 and in Fig 1.Linear algebra representation required to solve for the static distortion function f(x).

          The matrix on the left-hand-side, A, contains one row for each observed functional difference, and one column for each image pixel. Image sets of any dimensionality are handled by appending additional columns, detailed further below. In one dimension there are only two non-zero entries in each row, with coefficients of either +1 or -1 to implement the difference operator (e.g. f(x2) − f(x1) is shown in the first row). The vector f represents the unknown values of the distorting function to be solved. By multiplying the matrix A with the distortion map f, we would obtain the observed image registration data populated in the right-hand-side matrix Δ. Accordingly the unknown distortion map, f, can be solved by matrix inversion:

          alternatives f = A 1 Δ

          The array A is very large. If N is the number of pixels on the detector and M the number of image pairs considered (several images being recommended to ensure the solution is adequately constrained), there will be M*N observations (M*N rows) of the N variables (N columns). For most modern images this will be a very large array whose inverse would be prohibitively resource intensive to compute. However, it is not necessary to compute the inverse directly. Since the matrix is sparse there exists an efficient solution for the expression A-1 Δ through minimization of least squares. We implemented this using the Matlab function “lsqminnorm” (MATLAB R2017b, The Mathworks, USA). Although this function has additional benefits in selecting an appropriate solution to under-constrained problems, we found that it was necessary to constrain the problem by other means as detailed below.

          Constraining the solution

          With only two images as illustrated for simplicity above, the solution is poorly constrained. In a well-constrained solution, if two pixels are not directly related it would still be possible to learn the functional difference between them by tracing a series of paired functional differences across the image. Multiple image pairs are generally required to ensure that all pixels are relatable in this way to all other pixels, constraining the solution. When considering translations alone, typically several translations are required which should vary in degree and in direction in order to provide sufficient information. If these conditions are not met, translations alone can result in non-converging “streams” of pixel relationships across the image, such that some parts of the distorting function are not relatable to other parts.

          Interestingly, when rotation is included, this problem of non-relating “streams” dissolves because pixel relationships no longer travel in straight lines across the image. Extension of the method to handle any type of affine transformation, as opposed to just the translations illustrated above, is covered below.

          Handling non-integer displacements

          The simplified example above for feature “H” above (Table 1) considered direct 1:1 correspondences between the information seen by each pixel. In reality, precise registration will often yield fractional displacements, such that mapping is not 1:1 (e.g. in Table 1, “FG” was recorded at x = 1 in image #1).

          In principle this problem could be solved to arbitrary precision by upscaling the images, however this is impractical because the solution is already very resource intensive. Instead, we implemented a form of 2x2 anti-aliasing which produced satisfactory handling of fractional displacements. Under this scheme, the negative entries in the coefficient matrix, which were -1 in the simplified example, are distributed between the 4 pixels bounding the fractional pixel address, weighted by their proximity to that address. The sum of the coefficients for these 4 pixels remains -1, and the “reference” pixel is left unchanged at +1. The appropriate pixel weightings for each of the 4 bounding pixels were calculated by the product of independent weightings for x and y proximity, as follows:

          alternatives W x a , y b = 1 - | x - x a | * ( 1 - | y - y b | )

          The weightings calculated this way sum to unity for the 4 pixels, and dictate the proportion of the given object’s intensity that will be recorded in each of the 4 locations.

          Handling distortion in more than one spatial dimension

          The examples above refer only to displacements in the x direction. Real 2D or 3D image data can/will be distorted in the other dimensions as well. For each pixel (or voxel) in the image, a separate variable is needed for each dimension. For example in a 2D image with N pixels, one image pair will require 2N columns and 2N rows in the coefficient matrix A. Each additional image pair considered requires an additional 2N rows.

          In the case of pure translation, the array just described is unnecessarily large because the x and y (and z) problems could each be treated independently. The reason for structuring the array in this fashion is that in the case of other affine transformations of the scene, such as rotation, there is “cross talk” between the cardinal axes which must be accounted for, as described below.

          Affine transformation of co-ordinates

          The rationale given above concerns only translations between recorded image scenes. However, other affine transformations of the scene could be observed, with image rotation due to torsional movement of the eye being a known common problem in registration of ophthalmic imagery [35,36]. Affine transformations can be handled as follows:

          alternatives f x R f x M = x M T 1 , 1 * x R T 1 , 2 * y R T 1 , 3 f y R f y M = y M T 2 , 1 * x R T 2 , 2 * y R T 2 , 3

          Here the term T indicates a standard affine transformation matrix which can implement any combination of translation, rotation, scale and shear. The numbers in subscript indicate the row and column entries in the transformation matrix. For the example of rigid body transformations (translation and rotation alone), the transformation matrix (T) would appear as:

          alternatives T = c o s θ s i n θ t x s i n θ c o s θ t y 0 0 1

          Hence both the equations in x and in y have terms which account for the effect of rotations and translations. For example in the x-direction, it can be seen that in addition to accounting for a whole-field image shift T1,3 or tx, we must also take account of the rotation by transforming the found co-ordinates in the reference image (xR, yR). Importantly, the paired functional difference in the x-direction must now consider the y-location found in the reference image–there is interaction between the dimensions which must be accounted for as shown.

          It will be shown below that using the above transformations, rotation of the scene does not present any difficulty for the proposed algorithm. However, rotations must be estimated accurately for input to the algorithm; this can be a challenging task depending on the particular imaging modality [35,36]. Image registration is detailed in the next section.

          Image registration

          As with our previous work, the proposed algorithm should be conceptualised as a “meta-registration” procedure, meaning that it takes as input the results of successful image registration rather than carrying out the registration itself. Hence to demonstrate proof of principle it is not necessary to carry out “real” image registration; idealised values that were used to create the imagery were used. This saves computation time, but more importantly it avoids failure resulting from the image registration itself as opposed to a flaw in the proposed algorithm. Such failures would be hard to detect without manual review of the thousands of images simulated here.

          Nonetheless, it is important to demonstrate that the approach can be integrated with an image registration pipeline operating blindly on real data. Accordingly, we have implemented “real” registration in the example provided in S1 File. This involves initial estimation of the full-field shifts required to broadly register images in the stack, which was achieved with a standard Fourier domain cross-correlation approach. For estimation of distortion maps we employed the “demons” diffeomorphic registration algorithm [34], available in MATLAB through the Image Processing Toolbox (MATLAB R2017b, The Mathworks, USA). This algorithm was required rather than the “strip” registration conventionally employed for raster-scanned data, because static distortions will rarely be entirely directional so as to warrant a strip-based approach. The robustness of the proposed algorithm to errors in image registration may be explored by increasing the noise parameter included in S1 File.

          “Ground truth” data

          As in our previous work [10], “ground truth” images were obtained from by imaging the cone photoreceptor mosaic in a young, healthy subject at high speed (200 fps; frame exposure 2.5 ms) with a flood-illuminated adaptive optics ophthalmoscope and an imaging wavelength of 750 ± 25 nm. Images were acquired over a 7.5 mm pupil after adaptive optics had reduced the root-mean-square (RMS) wavefront error to 0.05 μm or less, i.e. below the diffraction limit. The ground truth data were montaged from multiple imaging locations temporal and within 2° of the foveal centre. This allowed a relatively wide field to be imaged without noticeable degradation in quality towards the edge of the field. Images of the photoreceptor mosaic in general provide an excellent example to evaluate the presence of image distortions due to the presence of high frequency information distributed uniformly across the image. In this specific dataset, there is negligible static or dynamic distortion because:

          • Images were acquired with a flood-illuminated system such that all pixels in a frame are acquired simultaneously. Dynamic distortions encountered across the field by raster sampling the moving eye are therefore avoided.
          • Image data presented are from a 512x512 pixel area which corresponds to approximately 0.9° in diameter. This is well within the lower bounds of the isoplanatic patch based on previously published estimates [18], hence, there should be minimal optical distortion encountered due to off-axis aberrations of the combined imaging system of the eye and ophthalmoscope.

          As in our previous work, the ground truth data were used here to simulate imagery acquired in the presence of distortion. The fidelity of distortion recovery was assessed by calculating the Pearson correlation in image intensity between the recovered and ground truth images. Applied distortion was either static alone, which could be produced by a “flood” illuminated system or by a scanning system operating at high frame rate or on a stationary eye; or a combination of static and dynamic distortion, which could be encountered with a scanning instrument in the presence of fixed sources of distortion. It should be noted that our simulation does not consider common differences which may be found between flood or scanning image modalities, for example regarding shot noise or speckle characteristics which are not expected to affect distortion per se.

          It should be noted that the algorithm described here operates on image registration data, rather than on the images themselves. As described above (see section “Image Registration”), for the majority of data presented here we carried out idealised registration does not consider the content of the images themselves; the images are merely useful tools to illustrate the degree of distortion. Hence, the use of cone photoreceptor images does not diminish the applicability of the proposed method to other types of images. The uniform, high spatial frequency content of such images does provide an excellent canvas in which to make comparisons, compared for example to images collected by other modalities which may lack “interesting” information across broad regions of the image, potentially masking residual distortion.

          This project was carried out in accordance with the principles expressed in the Declaration of Helsinki, and approved by the University of Melbourne Human Ethics Committee. Written informed consent was obtained from the subject prior to testing.

          Simulating full-field movements

          Simulating overall motion of the eye was accomplished by applying translations and rotations to the ground truth retinal scene, generally differing in degree and direction to provide complementary information as described above. Specific movements applied are detailed in the Results. This approach builds upon our previous work where only translations of the image were considered, and where data were limited to those eye movements which happened to occur during a particular imaging sequence.

          Since one cannot select what movements a real eye undergoes, we tested the new method with a large number of simulated runs, with the parameters for rigid body movement randomly drawn on each run. For this exercise the translation parameters were chosen from the range ± 10 pixels and the rotation parameter ± 5°. As described above under “Constraining the Solution”, it is necessary to ensure that images selected contain “unique” displacement information. We therefore applied somewhat arbitrary partial constraints to the translation of the retinal scene: translations each differed in magnitude by at least 2 pixels, and in direction from the origin by 45° or more. These constraints are by no means optimal, but could serve as an initial guide for selecting images from real data sets.

          Simulating distortion

          As described in the Introduction, it is difficult to know how much and what sort of static distortion is present in ophthalmic imaging systems. Our general philosophy was therefore to simulate distortion of greater magnitude and with more complex structure across the field than we believe might be encountered in practice. This should serve to convince the reader of the robustness of the proposed method.

          To test static distorting functions of relatively arbitrary shapes, we applied 2D sine functions (in x and y) of varying amplitude, phase and frequency as follows, where x and y ranged from 0 to 2π over the full image array:

          alternatives D x x , y = A x sin f x x φ x D y x , y = A y sin f y y φ y

          There were accordingly 6 parameters set for each distorting function. It should be noted that the horizontal distortion in Eq 5 varies only in x, and the vertical distortion only in y. This makes it easier to visualize the results because the distortion map can be meaningfully plotted in one dimension. However the method is not limited to application in this way and works just as well for arbitrary distortions. Similarly, the method works just as well for simpler distortions, for example the rotationally symmetric distortions expected for a centred, high power system. We chose to present results from sinusoidal distortions here, in part because they look sufficiently bizarre to convince the reader of the broad applicability of the method, and in part because of the well-known Fourier decomposition theorem which allows arbitrary shapes to be reconstructed from a sinusoidal series.

          The above parameters allowed us to test distortions of high amplitude, and with unusual structure (e.g. asymmetric, minimum not coinciding with the array centre, multiple cycles across the image) to ensure robustness of our approach. In the Results presented below, the amplitude parameters lay in the range ± 5 pixels, the frequency parameter in the range 0 to 5 cycles, and the phase parameter in the range 0 to 2π.

          We also present data in which dynamic and static distortions are combined. Dynamic distortions were implemented by superimposing full-field movement of the scene with 2D sinusoidal distortion following Eq 5, with new free parameters randomly drawn for each frame. To mimic the behaviour of raster systems whereby minimal dynamic distortion is expected along the “fast” scan axis, for dynamic distortion we modelled variation of the distorting functions only in the y direction (i.e. Dx(y) and Dy(y)). This assumption could be violated during very fast (saccadic) eye movements, however, we do not consider such movements here due to the irrecoverable motion blur which would result.

          Although the sinusoidal distortions modelled do not have particular “real life” analogues (the eye is unlikely to follow sinusoidal motion), this approach allowed simulation of distorting functions that varied widely in degree and spatial structure in an unbiased manner. There is also no correlation in distortion applied between one frame and the next as would occur for a real eye. This helps to avoid a potential criticism of our previous work [10], where it might be supposed that the eye movement data underlying our simulations happened to be a fortuitous set for the purposes of image recovery. We note further that our previous work [10] and its subsequent application by others [1217] have already confirmed the dynamic correction method to correct for dynamic distortion in the presence of realistic eye movement patterns. In the modelling below, we will show that as long as a satisfactory dynamic correction is obtained, the new static correction procedure can then be applied to recover images that are free of both kinds of distortion. The significant distortion entailed by the present modelling can be appreciated by the sequence of 100 consecutive frames shown in S1 Fig.

          Dynamic correction method

          Although described in full elsewhere, it is worthwhile to give a brief overview of our previously published method for correcting dynamic distortions [10]. These are generated dynamically in raster-based imagery because eye movements occur during the sequential acquisition of image points. The method relies on the assertion that, given the stochastic nature of fixational eye movements, the expected value of distortion “seen” by one part of the retina should not differ from that expected for any other part of the retina. In other words, the bias in apparent displacement for each piece of tissue should trend towards zero when many frames are considered. Accordingly, the local distortions accrued within any particular frame can be discovered by registering each piece of tissue therein across a large number of frames; if it is found that the average displacement for that tissue differs from zero, this must result from the distortion inherent in the frame in question, such that the net bias does sum to zero.

          The assumption of zero bias described above must be violated in the case of systematic error which produces static distortion, present on every frame. In that case, assuming that the bias is zero should be expected to result in the full degree of underlying systematic bias remaining within the “recovered” frame. This underpins the strategy advanced in the current work, whereby dynamic recovery is first applied in order to produce a handful of frames which all suffer from a common systematic bias.

          Array size

          Similar to our previous publication, initial dimensions of simulated images were 400 x 400. However, the high memory requirements of the proposed algorithm left us unable to solve arrays of this size on a laptop computer with 16 GB RAM. We therefore clipped to the central 200 x 200 when supplying images and distortion maps to the DIOS algorithm. This limitation and potential solutions to it are expanded on further in the Discussion.

          Results

          Static distortion only

          Fig 3 shows example scenes of the cone photoreceptor mosaic with the eye in slightly different positions and with 2D static distortion applied (i.e. the warping function is identical in each frame). The true distorting functions in x and y are plotted in Fig 3). Four frames were simulated with mean position given by the coloured crosses; two example frames are shown (Fig 3A and 3B). There was no rotation between images in this example. Following registration of the images to the first frame using the “demons” algorithm as described above, the registration information was passed to the DIOS algorithm. Example images recovered are shown in Fig 3D and 3E, giving strong correspondence to the ground truth (Fig 3C). Similarly the recovered distortion function is very close to the function that was used to generate the images (Fig 3F).

          Example correction of static distortion in 2D.
          Fig 3
          Four distorted frames were generated with the eye at different positions corresponding to the coloured crosses. Example imagery is shown (A, B) which appears highly warped compared to the ground truth data (C). Nonetheless, images are recovered with high accuracy (D, E). The profile of recovered distorting functions is plotted (F), showing strong similarity to the true distorting profiles.Example correction of static distortion in 2D.

          Fig 4 illustrates two potential pitfalls of the proposed algorithm. The same warped examples as Fig 3 are shown, however now only two of the frames were input, which are offset from one another in the vertical direction only (Fig 4C, crosses). It can be seen that strong distortion remains in the x direction (Fig 4D and 4E) and that a flat distortion profile was estimated in this direction (Fig 4F). This demonstrates that images need to be offset from one another obliquely if they are to facilitate recovery of image distortion in 2D (and displacements would presumably be required in the third dimension as well in the case of 3D data). For this reason we have termed the approach “DIOS”, or Dewarp Image by Oblique Shift. The other pitfall illustrated here is seen in Fig 4F, where even in the y-direction “wiggles” are apparent in the shape of the recovered distorting function. This illustrates that the solution is poorly constrained, which occurs because only a single frame pair was input to the procedure. As described above this leads to “streams” of pixel relationships across the image, such that certain pixels cannot be related to other pixels in the image.

          Example failure to correct for static distortion in 2D.
          Fig 4
          Two distorted frames were generated with the eye at different positions corresponding to the coloured crosses. Generated images (A, B) and ground truth (C) correspond to those in Fig 1. Recovered images are still seen to be highly warped in the horizontal direction (D, E). This occurs because there is no information provided regarding differences in the horizontal direction (flatline in F). In the vertical direction, the use of just two frames fails to constrain the solution, producing a “shakey” appearance to the recovered distortion profile (red line in F).Example failure to correct for static distortion in 2D.

          Fig 5 demonstrates that the proposed procedure can properly handle rotations of the scene, that is, torsional movements of the eye. Fig 5 illustrates data simulated in the same way as Fig 3, but with rotations spanning ± 30°. It can be seen that, as for Fig 1, recovered images and distorting functions appear very similar to the ground truth, indicating that rotation was adequately addressed.

          Example correction in the presence of rotation.
          Fig 5
          Four distorted frames were generated with the eye at positions as for Fig 1, however now with rotation spanning 50° between the 4 frames. Example images are shown (A, B) which are rotated by 30° with respect to each other. Image A is the same as that for Fig 3. Both example images are recovered with high accuracy (D, E) and recovered distorting profiles (F) match the ground truth, demonstrating robustness to image rotation.Example correction in the presence of rotation.

          To demonstrate that the method can handle distortions of arbitrary form, we modelled 1,000 static distorting functions whose parameters (Eq 5) were drawn uniformly from the full range described above (see section “Simulating distortion”). Given the computation requirements for larger arrays, here we recovered only the central 100x100 pixel region. The simulation parameters supported extreme distortions (the most distorted example is shown in Fig 6A). For each simulated run shown in Fig 6, we “acquired” 4 frames with the eye having been translated and rotated by random amounts (see section “Simulating full-field movements”).

          Outcomes from 1000 different static distortion maps.
          Fig 6
          On each run a random 2D distorting function was generated, and sampled at 4 randomly selected gaze positions (see text for details). A) Shows the worst simulated image, i.e. the one least correlated with the ground truth. B) Shows the recovered image for A. C) Plots the baseline (red) and recovered (blue) correlation with the ground truth for all 1000 runs. Only 17/1000 recovered images failed to reach a correlation above 0.99.Outcomes from 1000 different static distortion maps.

          The outcomes of these 1,000 simulated runs are plotted in Fig 6. Even for the worst case simulated (Fig 6A), the recovered image was very similar to the ground truth (Fig 6B; correlation > 0.99). Fig 6C shows the initial (red) and recovered (blue) correlation for all 1,000 simulated runs, confirming excellent similarity to the ground truth image over the vast majority of simulated runs (only 17/1000 runs showed correlation below 0.99).

          Static and dynamic distortion

          The DIOS procedure may be applied in sequence with the dynamic recovery procedure previously described [10] to recover images that are, in principle at least, distortion-free. To illustrate this process we modelled combined dynamic and static distortion, with static distortion as shown in Fig 7 (black). This is a destructive example because all static distortion coincides in direction with that of the dynamic distortion (the direction of the “slow” scan axis). Each frame was also warped dynamically, with distortion in both x and y directions generated from Eq 5 with frequency and phase parameters drawn from the full range described above (see Methods section “Simulating distortion”). The amplitude of the distorting function here was fixed at 5 pixels to ensure dynamic warp of all frames (i.e. contrary to real data, there is no single frame which could match the ground truth). Example frames warped by both static and dynamic distortion are shown in Fig 8A and 8B.

          Static distortion profiles applied and recovered in the presence of dynamic distortion.
          Fig 7
          Six clusters of 1000 frames were generated (see text and Fig 8). Average distortion along the y array direction is plotted for vertical distortions (A) and horizontal distortions (B). The recovered solution is shown in cyan, matching the required solution in blue. The true underlying static distortion is shown in magenta, but this is only the correct solution when the eye did not move vertically (red). Black shows the solution when there were only eye movements (no sinusoidal warp applied).Static distortion profiles applied and recovered in the presence of dynamic distortion.
          Correction of simultaneous static and dynamic distortion.
          Fig 8
          A) Example image from a cluster of 1,000 frames, warped by both static and dynamic distortion. B) Example image from a second cluster; only dynamic distortion differs from A. C) Ground truth image, overlaid with the gaze position for six 1,000 frame clusters (each cluster a different colour). D) Output of the dynamic recovery procedure on the cluster corresponding to A. E) Output of the dynamic recovery procedure on the cluster corresponding to B. F) Output of the DIOS solution, running on the six outputs of dynamic recovery (one for each cluster), examples of which were shown in D and E. Red polygons: a constellation of cells to aid visualization of correction for distortion. Yellow triangles: the same constellation of cells in the ground truth image, shifted and rotated as needed to minimise error.Correction of simultaneous static and dynamic distortion.

          Distorted frames were generated in 6 “clusters”, with each cluster having slightly different mean gaze position (up to 10 pixels) as required by the DIOS algorithm. Each cluster also had a different torsional component (up to 10°), to demonstrate robustness to image rotations. Clusters consisted of 1000 frames to ensure that bias resulting from dynamic warp would trend close to zero. Unlike the simulation of random static distortions that was shown in Fig 6, there was no special requirement imposed on gaze position within a cluster, other than that it be randomly drawn from a normal distribution (scaled so that max. amplitude within a cluster was 50 pixels). The simulated eye movement pattern within each cluster was independent from the other clusters. The eye position for all frames and all clusters is indicated in Fig 8C (coloured dots; a different colour for each cluster), overlaid on the ground truth image.

          The dynamic recovery procedure was run on each cluster independently, producing a single recovered frame (examples shown in Fig 8D and 8E). These frames appear qualitatively to be much less distorted than the example input frames (Fig 8A and 8B), however significant distortion remains (e.g. Fig 8D has a correlation of 0.67 to the ground truth image). This can be appreciated visually by comparing an example constellation of 5 cells (red polygons) with the same constellation from the ground truth image (yellow polygons). Applying dynamic recovery to all 6 clusters provided 6 frames for input to the DIOS procedure, which returned the solution plotted in Fig 7 (cyan). The distortion required to faithfully render the ground truth image without error is shown in blue; their strong overlap indicates that the solution was accurate (RMS error = 0.18 pixels). This can be further verified by the similarity with the ground truth image of Fig 8C (correlation = 0.991; recovering the images drawn from other clusters gave near-identical results with the worst correlation at 0.986) and by the reference constellation (overlap of yellow and red polygons in Fig 8F).

          Interestingly, in this example the required correction differed substantially from the underlying static distortion (compare cyan and magenta curves in Fig 7A). This means that the bias remaining after dynamic recovery was not simply equivalent to the underlying static distortion, indicating some interaction effect. To investigate this, we zeroed the vertical component of within-cluster eye movements and repeated the simulation. The solution obtained is shown by the red curve in Fig 7, which now closely resembles the underlying static distortion in magenta (RMS error = 0.19 pixels). This result suggests that panning the tissue across different regions of the static distorting function conferred a pseudo-dynamic distortion. Despite this, the combination of the dynamic and DIOS procedures seemed able to recover the solution as was shown in Fig 8. To learn whether the dynamic, sinusoidal distortion also interacted with the static component, we repeated the simulation with zero dynamic distortion (and with the original eye movements). This resulted in the black curve of Fig 7, which is highly similar to the underlying function that was applied, indicating minimal interaction (RMS = 0.17 pixels).

          Discussion

          We have demonstrated a new method to estimate and correct for static distortions in an imaging system. The method is particularly suited for ophthalmic imaging because: a) no calibration target is required, which is not available in the eye; b) no model is assumed for the distorting function; c) the method requires only full-field movement of the scene which is guaranteed by the incessant motion of the eye; d) the method can handle torsional movements which can occur for the fixating eye, but does not rely upon them; e) the method can be used in series with a previously published approach to correct dynamic distortion as well, which is encountered by modern ophthalmic imaging modalities.

          The method takes as input the pixelwise registration maps between a series of images and a reference image, and assumes that the affine displacement (e.g. translation, rotation) between images can be accurately inferred despite the presence of image warp. This affine information is used to “correct” the registration maps, such that it provides information on the difference in the distorting function between pairs of pixels on the detector. If multiple images are acquired with the eye having moved in different directions by different amounts, this series of observations can be used to solve for the relative shape of the distorting function. The method is blind to the absolute value of the distorting function, that is, to a fixed prismatic shift applied to all images. Such shifts merely move the image rather than producing distortion. The method would also be blind to uniform scaling (magnification) of the entire image set, unless the true amount of eye movement is known.

          Our modelling suggests that the method is capable of arbitrary precision, with residual errors a small fraction of a pixel. However, there are some caveats to this. A sufficiently large number of frames (more than two) must be acquired, with the eye having moved in different directions by different amounts, in order to adequately constrain the solution. Where this is not the case, there are effectively “islands” of connected pixels produced within the image. A simple example where this occurs would be when the translation between images is vertical only, as in Fig 4; the “match” for each pixel will then always be found above or below it, but never to the right or left. This means that each column of the image effectively becomes an independent system of equations to solve. Because an arbitrary offset must be set for each system of equations (as the algorithm only solves for relative differences), these islands will invariably end up at different absolute displacement values, producing a “bumpy” appearance to the solution. One limitation at present is that we can offer no guidance as to whether a particular set of affine transformations will provide sufficiently unique information to constrain the solution; the output needs to be reviewed for consistency.

          Any correction applied to real images must use real image registration, which will be subject to some degree of error. Although the least squares nature of the solution is likely to be resilient to such errors, we have explored this only briefly (see S1 File), wishing here primarily to demonstrate in principle that the method itself does not impose any significant limitation to accuracy. Note that consideration of error due to inaccurate image registration encompasses errors which may arise due to frame-to-frame variations in intensity. Such variations could be caused by factors such as light-evoked fluctuations in reflectance, the oxygenation state of blood, shot noise, non-uniformity of illumination, vignetting, etc. Where these do not cause error in image registration, such intensity variations will not affect the quality of the recovered distortion maps, because the algorithm presented operates only with registration information rather than intensity information.

          To correct for both simultaneous static and dynamic distortion, one strategy is to first correct for dynamic distortion within a given “cluster” of frames, and to pass the output from several such clusters to the DIOS procedure. The mean position of the eye for each cluster needs to be somewhat different in order to provide the required offsets for the DIOS algorithm; this could be achieved in a real dataset by judicious separation of a full set of frames, acquired at the same nominal position of fixation, into pseudo-clusters that each carry the desired mean position. There is no requirement that the clusters need be composed of entirely different frames or that they be contiguous in time, and differences in mean position required need only be a handful of pixels (<= 10 in our simulations). In our modelling we used a very large number of frames in each cluster to meet the fundamental assumption of the dynamic recovery procedure (i.e. that bias in displacement for each particular piece of anatomy is close to zero). Where this assumption is not met (e.g. if too few frames are considered), artifactual distortion could remain that will not be common with other clusters of frames, which could produce error in the DIOS solution.

          It may at first appear surprising that the output of the previously described dynamic correction procedure would combine well with the newly proposed static correction procedure. However, this marriage is assured by the large number of frames that we included within each cluster. This means that the interaction occurring between systematic and dynamic warp within one cluster should be expected to be nearly identical to any such interaction taking place in another cluster. In other words, the residual static error should not differ appreciably between clusters, given the large number of frames within each cluster. This satisfies the conditions for successful application of the DIOS procedure (a distortion function that is common to all clusters), such that near-perfect correction should be expected. It was noteworthy, though, that the residual static correction obtained was not simply equal to the ground truth static correction applied. Further work should explore the interaction between systematic and dynamic sources of distortion, which is likely to depend on the degree and spatial frequency of each type of distortion.

          One alternative to the use of clusters with very many frames may be to use a handful of single frames where the eye has undergone ostensibly the same pattern of movement within each frame. Where this occurs, dynamic distortion will be common between each image. Together with the common static distortion, the image could be considered to suffer from common pseudo-static distortion that can then be corrected by the DIOS procedure alone (i.e. without first requiring the dynamic recovery procedure). We have confirmed this to work in simulations (data not shown), however the question remains as to how many frames one would need to acquire in order to possess a handful with sufficiently similar eye movements to support this approach.

          A limitation of the approaches advanced above to correct for static and dynamic distortion is that one may not be able to perform a “once-off” system calibration, or to solve for eye movements with complete accuracy, due to interactions between static distortion and dynamic sources of distortion (e.g. the difference between red/magenta and the other curves in Fig 7, showing respectively that the underlying static distortion is not necessarily equal to the apparent or required static distortion). Although, we note that this may no longer be true if static and dynamic distortions are orthogonal; for example in the SLO desinusoiding error is produced in the horizontal direction, whereas eye movements result in distorting functions that change only in the vertical direction.

          Some of the limitations discussed above may be overly pessimistic and not detrimental in practice. In the field of AO-SLO imaging, any static distortions are expected to be less obvious than modelled here (e.g. worst case in Fig 6A). In addition, most groups employ various strategies to identify frames which appear minimally distorted for the purpose of providing a reference frame for image registration; if similar heuristics are applied to discard overly distorted frames before attempting the dynamic recovery procedure, residual errors are likely to be relatively low compared to the modelling undertaken here where all frames were included.

          For retinal imaging systems adopting a wider field of view than a typical adaptive optics design, distortion is likely to be far greater as discussed in the Introduction. As noted in that section, there is a lack of experimental evidence regarding the degree of distortion imposed by the eye in conjunction with typical imaging systems. However, our method can be used to redress this lack of evidence, being applicable to data acquired from any ophthalmic imaging device. If the degree of distortion calculated is found to be substantial, the proposed algorithm can then be implemented in the image processing software of the particular device to correct image data saved in the future, and can also be applied post hoc to correct previously acquired data if multiple frames were recorded.

          One limitation that is likely to be encountered in practice is the processing resources required for the algorithm. As described above, we solved for a 200x200 pixel patch of the image array rather than the full array. Some clipping was warranted to avoid areas of missing information due to eye movements, however, the major reason for using the smaller region of interest was that larger array sizes were not solvable on a system running with 16 GB RAM. Processing time was also several minutes for the 200x200 patch (not including image registration). Some of these issues should be addressable through more refined implementation of the algorithm, e.g. the least squares solution should be parallelisable, which should improve processing time by use of GPU computing, and also permit the use of hard disk storage to supplement RAM requirements. Where there are no rotations present, x and y solutions can be derived independently which also reduces memory requirements. However it should be noted that, in the presence of rotation, the least squares minimisation function typically converged more quickly on its solution.

          In the meantime, a more pragmatic solution could be to divide the detector up into “tiles”, which would be made to partially overlap to appropriately anchor the absolute offset applied. Another option to reduce computing requirements is to simply bin the registration data that is input to the algorithm, since typical distorting functions are unlikely to change to a significant degree between adjacent pixels. The solution obtained could then be upsampled and applied to the full-sized array. This solution is demonstrated in the example supplied in S1 File. Another option still could be a pyramid approach, where lower frequency information is solved first, from sub-sampled data, before turning to progressively higher frequency information sampled at greater resolution.

          It should be noted that the general nature of our simulations suggest that the proposed method could be applied to imaging systems of arbitrary type, not just the eye. As long as it is possible to perform diffeomorphic registration between acquired images, the method should be applicable to any imaging system which suffers from static distortion and where it is possible (or, as in the eye, unavoidable) to pan the scene obliquely across the image array.

          Finally, the method advanced here has application beyond simple imaging, to understanding the evolution of visual systems and machine vision. It is intriguing to wonder whether the small, incessant movements made by the fixating eye may have evolved, in part, due to their utility in calibrating the visual scene. This would be a useful ability for an eye to develop due to its high optical power and non-regular arrangement of receptor elements. Previous work has described how comparison of the visual scene after small movements of the eye/head might be used to iteratively arrive at a satisfactory calibration for such causes of distortion [33]. The present work extends such thinking, providing an analytical solution that can be solved with only a handful of samples of the visual scene. We have also shown that the method can be augmented to handle temporal variations, which could arise due to non-synchronous firing of neurons or due to random movement within the scene, as occurs for example by shimmer arising from the ground on a hot day.

          References

          Kashani AH , Chen C-L , Gahm JK , Zheng F , Richter GM , Rosenfeld PJ , et al . Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications. Progress in retinal and eye research. 2017;60:66100. doi: doi: 10.1016/j.preteyeres.2017.07.002

          Sakata LM , DeLeon‐Ortega J , Sakata V , Girkin CA . Optical coherence tomography of the retina and optic nerve–a review. Clinical & experimental ophthalmology. 2009;37(1):909. doi: doi: 10.1111/j.1442-9071.2009.02015.x

          Bennett TJ , Barry CJ . Ophthalmic imaging today: an ophthalmic photographer’s viewpoint–a review. Clinical & experimental ophthalmology. 2009;37(1):213.

          Bedggood P , Metha A . Recovering the appearance of the capillary blood column from under-sampled flow data. Optics Letters. 2020;45(15):43203. doi: doi: 10.1364/OL.398168

          Burns SA , Elsner AE , Sapoznik KA , Warner RL , Gast TJ . Adaptive optics imaging of the human retina. Progress in retinal and eye research. 2019;68:130. doi: doi: 10.1016/j.preteyeres.2018.08.002

          Jonnal RS , Kocaoglu OP , Zawadzki RJ , Liu Z , Miller DT , Werner JS . A review of adaptive optics optical coherence tomography: technical advances, scientific applications, and the future. Investigative ophthalmology & visual science. 2016;57(9):OCT51OCT68.

          Carpenter RH. Movements of the Eyes, 2nd Rev: Pion Limited; 1988.

          Vienola KV , Braaf B , Sheehy CK , Yang Q , Tiruveedhula P , Arathorn DW , et al . Real-time eye motion compensation for OCT imaging with tracking SLO. Biomedical optics express. 2012;3(11):295063. doi: doi: 10.1364/BOE.3.002950

          Salmon AE , Cooper RF , Langlo CS , Baghaie A , Dubra A , Carroll J . An automated reference frame selection (ARFS) algorithm for cone imaging with adaptive optics scanning light ophthalmoscopy. Translational vision science & technology. 2017;6(2):9. doi: doi: 10.1167/tvst.6.2.9

          10 

          Bedggood P , Metha A . De-warping of images and improved eye tracking for the scanning laser ophthalmoscope. PloS one. 2017;12(4):e0174617. doi: doi: 10.1371/journal.pone.0174617

          11 

          Vogel CR , Arathorn DW , Roorda A , Parker A . Retinal motion estimation in adaptive optics scanning laser ophthalmoscopy. Optics express. 2006;14(2):48797. doi: doi: 10.1364/opex.14.000487

          12 

          Azimipour M , Zawadzki RJ , Gorczynska I , Migacz J , Werner JS , Jonnal RS . Intraframe motion correction for raster-scanned adaptive optics images using strip-based cross-correlation lag biases. PloS one. 2018;13(10):e0206052. doi: doi: 10.1371/journal.pone.0206052

          13 

          Li Z , Pandiyan VP , Maloney-Bertelli A , Jiang X , Li X , Sabesan R . Correcting intra-volume distortion for AO-OCT using 3D correlation based registration. Optics Express. 2020;28(25):38390409. doi: doi: 10.1364/OE.410374

          14 

          Cooper RF , Aguirre GK , Morgan JI . Fully Automated Estimation of Spacing and Density for Retinal Mosaics. Translational vision science & technology. 2019;8(5):26. doi: doi: 10.1167/tvst.8.5.26

          15 

          Bowers NR , Boehm AE , Roorda A . The effects of fixational tremor on the retinal image. Journal of vision. 2019;19(11):8. doi: doi: 10.1167/19.11.8

          16 

          Luo T , Gast TJ , Vermeer TJ , Burns SA . Retinal vascular branching in healthy and diabetic subjects. Investigative Ophthalmology & Visual Science. 2017;58(5):268594.

          17 

          Mecê P , Scholler J , Groux K , Boccara C . High-resolution in-vivo human retinal imaging using full-field OCT with optical stabilization of axial motion. Biomedical optics express. 2020;11(1):492504. doi: doi: 10.1364/BOE.381398

          18 

          Bedggood P , Daaboul M , Ashman RA , Smith GG , Metha A . Characteristics of the human isoplanatic patch and implications for adaptive optics retinal imaging. Journal of biomedical optics. 2008;13(2):024008. doi: doi: 10.1117/1.2907211

          19 

          Bedggood P , Metha A . System design considerations to improve isoplanatism for adaptive optics retinal imaging. JOSA A. 2010;27(11):A37A47. doi: doi: 10.1364/JOSAA.27.000A37

          20 

          Dubra A , Sulai Y . Reflective afocal broadband adaptive optics scanning ophthalmoscope. Biomedical optics express. 2011;2(6):175768. doi: doi: 10.1364/BOE.2.001757

          21 

          Akondi V , Kowalski B , Burns SA , Dubra A . Dynamic distortion in resonant galvanometric optical scanners. Optica. 2020;7(11):150613.

          22 

          Escudero-Sanz I , Navarro R . Off-axis aberrations of a wide-angle schematic eye model. JOSA A. 1999;16(8):188191. doi: doi: 10.1364/josaa.16.001881

          23 

          Bedggood P , Ashman R , Smith G , Metha A . Multiconjugate adaptive optics applied to an anatomically accurate human eye model. Optics express. 2006;14(18):801930. doi: doi: 10.1364/oe.14.008019

          24 

          Friberg TR , Pandya A , Eller AW . Non-mydriatic panoramic fundus imaging using a non-contact scanning laser-based system. Ophthalmic Surgery, Lasers and Imaging Retina. 2003;34(6):48897.

          25 

          Zhang Z . A flexible new technique for camera calibration. IEEE Transactions on pattern analysis and machine intelligence. 2000;22(11):13304.

          26 

          Zhang Z . Camera calibration: a personal retrospective. Springer; 2016.

          27 

          Smith G , Bedggood P , Ashman R , Daaboul M , Metha A . Exploring ocular aberrations with a schematic human eye model. Optometry and Vision Science. 2008;85(5):33040. doi: doi: 10.1097/OPX.0b013e31816c4449

          28 

          Nistér D, Stewénius H, Grossmann E, editors. Non-parametric self-calibration. Tenth IEEE International Conference on Computer Vision (ICCV’05) Volume 1; 2005: IEEE.

          29 

          Grossmann E, Lee E-J, Hislop P, Nister D, Stewenius H, editors. Are two rotational flows sufficient to calibrate a smooth non-parametric sensor? 2006 IEEE Computer Society Conference on Computer Vision and Pattern Recognition (CVPR’06); 2006: IEEE.

          30 

          Ramalingam S , Sturm P , Lodha SK . Generic self-calibration of central cameras. Computer Vision and Image Understanding. 2010;114(2):2109.

          31 

          Von Gioi RG, Monasse P, Morel J-M, Tang Z, editors. Towards high-precision lens distortion correction. 2010 IEEE International Conference on Image Processing; 2010: IEEE.

          32 

          Espuny F , Gil JIB . Generic self-calibration of central cameras from two rotational flows. International Journal of Computer Vision. 2011;91(2):13145.

          33 

          Maloney LT , Ahumada AJ . Learning by assertion: Two methods for calibrating a linear visual system. Neural Computation. 1989;1(3):392401.

          34 

          Vercauteren T , Pennec X , Perchant A , Ayache N . Diffeomorphic demons: Efficient non-parametric image registration. NeuroImage. 2009;45(1):S61S72. doi: doi: 10.1016/j.neuroimage.2008.10.040

          35 

          Yang Q , Zhang J , Nozato K , Saito K , Williams DR , Roorda A , et al . Closed-loop optical stabilization and digital image registration in adaptive optics scanning light ophthalmoscopy. Biomedical optics express. 2014;5(9):317491. doi: doi: 10.1364/BOE.5.003174

          36 

          Kurokawa K , Crowell JA , Do N , Lee JJ , Miller DT . Multi-reference global registration of individual A-lines in adaptive optics optical coherence tomography retinal images. Journal of Biomedical Optics. 2021;26(1):016001.

          3 Feb 2021

          PONE-D-20-39508

          Towards distortion-free imaging for the moving eye

          PLOS ONE

          Dear Dr. Bedggood,

          Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

          Please submit your revised manuscript by March 1, 2021. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

          Please include the following items when submitting your revised manuscript:

          • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
          • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
          • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

          If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

          If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

          We look forward to receiving your revised manuscript.

          Kind regards,

          Yuhua Zhang

          Academic Editor

          PLOS ONE

          Journal Requirements:

          When submitting your revision, we need you to address these additional requirements.

          1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

          https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

          https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

          2. If materials, methods, and protocols are well established, authors may cite articles where those protocols are described in detail (for instance here, Bedggood P, Metha A. De-warping of images and improved eye tracking for the scanning laser ophthalmoscope. PloS one. 2017;12(4):e0174617), but the submission should include sufficient information to be understood independent of these references

          (https://journals.plos.org/plosone/s/submission-guidelines#loc-materials-and-methods)

          [Note: HTML markup is below. Please do not edit.]

          Reviewers' comments:

          Reviewer's Responses to Questions

          Comments to the Author

          1. Is the manuscript technically sound, and do the data support the conclusions?

          The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

          Reviewer #1: Partly

          Reviewer #2: Yes

          **********

          2. Has the statistical analysis been performed appropriately and rigorously?

          Reviewer #1: N/A

          Reviewer #2: No

          **********

          3. Have the authors made all data underlying the findings in their manuscript fully available?

          The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

          Reviewer #1: No

          Reviewer #2: Yes

          **********

          4. Is the manuscript presented in an intelligible fashion and written in standard English?

          PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

          Reviewer #1: Yes

          Reviewer #2: Yes

          **********

          5. Review Comments to the Author

          Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

          Reviewer #1: Review of PONE-D-20-39508

          The authors demonstrated a concept of a method for correcting systematic distortions in scanning-based ophthalmic imaging instruments. Although only corrections of simulated distortions are presented, the concept itself might be an interest of the field of ophthalmic imaging.

          However, the principle of the method is not well written, and all simulation parameters are not well organized. There are some questions in the result of correction for systematic and dynamic distortions (LINE 418-479). These points are required to be addressed before publication.

          - Classification of distortions

          The word "systematic distortion" sounds to me that static distortions due to an imaging instrument. In fact, the authors use "dynamic distortions" as a counterpart. But then the "pseudo-systemic distortion" (LINE 27) seems odd. Because "DIOS" and "dynamic restoration methods" are image-based heuristics, introducing another classification based on the type of image distortions, such as "static image distortions" and "dynamic image distortions," may help clarify the discussion of results.

          -Methods

          --Characteristics of imaging modalities

          The authors did the proof of concept of their idea, DIOS. The images without distortions are used as ground truth, and scanning ophthalmoscope images are simulated by distorting them. To obtain the "original" ground truth images, the authors used a flood-illumination AO ophthalmoscope.

          The main concern is whether the different imaging characteristics between flood-illumination-based and scanning-based ophthalmoscopes affect the performance of the presented method.

          For example, speckle appearance might be different between them. The images of the flood illumination AO ophthalmoscope exhibit less speckle. However, an image of single frame of scanning ophthalmoscope may express more speckle.

          For the readers to properly understand the possible difficulties, these points should be discussed, and the details of the system should be described, although the previous authors' paper [10] is cited.

          --DIOS principle

          Overall the principle of DIOS is quite difficult to understand. Equations are not well written, and there is no consistency among equations, figures, and tables.

          --- Line 96

          What does the symbol "T" mean? Although the description appears later (Line 136), it should be explained where the symbol first appears.

          ---Line 117

          What does mean "the reference image (xR)" and "the moving image, xM"? According to the description around Equation 1, it might be the locations of the same object in the reference image and the moving image.

          ---Line 119-120

          The numbers (2, 1, 0, 3, -3) suddenly appear. Please explain the details. Do the authors use "image 1" as "Reference image" and "image 2" as "Moving image"?

          ---Line 124

          From two formulae ("Reference image" and "Moving image"), Equation 1 cannot be derived. The symbol "T" should disappear. Perhaps the authors should discriminate it for the reference image and moving image as "TR" and "TM".

          ---Figure 1

          Figure 1 is unclear. What is the vertical axis? What do the labels "f(x1)", "f(x2)", and "f(x3)" mean? What do the circles mean?

          ---Line 215

          The description of "anti-aliasing" is too short. Please describe more details, or please cite literature if there are.

          The reason why the anti-aliasing can achieve a precision similar to upscaling is unclear.

          ---Equation

          Where is Equation 5?

          --Line 312

          What are 2D sine functions? Equation 6 describes two 1D sine functions. Do the authors mean that both Dx and Dy are distributed in the xy-plane? If it is, Equation 6 should be corrected.

          -Results

          --Systematic distortion, dynamic distortion, and eye movements

          There are many parameters in simulations of distortions and eye movements. It is difficult to follow the logic and interpret the results.

          For each result, all simulation parameters and constraints must be explicitly described; otherwise, the results cannot be interpreted correctly. There are many ambiguities in the current manuscript. For example, the constraints to translation (eye movements simulation?) described from the Line 400 to Line 404 are commonly applied to all results?

          In my opinion, the parameters and constraints of the simulation should be listed in the "Methods" section since they are the parameters of the "Verification Method."

          --Systematic and dynamic distortion

          As shown in Figs. 7 and 8, and discussed by the authors, there will be the interaction between distortion and eye movement. The authors conclude that the combination of the presented method (DIOS) and the previous method (dynamic recovery procedure) correct the distortions well.

          This is a surprising result because the principle of DIOS assumes a common distortion (f) among image pairs. Did the authors apply common within-cluster eye movements for all clusters? Why is only one image (Fig. 8F) presented? What about other 5 images after the DIOS procedure?

          If the authors' insight is correct, the discrepancy of the solution from the set systematic distortion is coming from the interaction between the distortion and eye movements. Then, the solution for correcting distortions depends on eye movement.

          In order to support their conclusions, the authors need to simulate independent random intra-cluster eye movements and compare the similarity between the final corrected images with DIOS derived from each cluster.

          Reviewer #2: This paper describes a method for correction of static distortion in an imaging system. This was a surprise to me, after having read only the title and the first sentence of the abstract, which led me to believe that the paper would treat the dynamic distortion caused by the interaction of intra-frame motion of the eye with a scanning system. In fact, this paper does not address that particular problem at all. For that reason, I would retitle the paper, and rewrite the abstract, possibly eliminating mention of dynamic distortion in the abstract, but in any case not leading with it!

          It is really not clear to me that the method as presented has anything to do with the eye or ocular imaging at all. Clearly the authors hope to apply it to that domain, and have used high-magnification retinal images as their test images in their simulations. But the synthetic distortions that are corrected are arbitrary sinusoidal warps, that to my knowledge are common distortions in any sort of imaging system, ocular or otherwise.

          Having reread the abstract, it is not obvious that I have fully understood the paper, as the sentence "the method successfully handles torsional movement of the eye" seems to go beyond correction of a static distortion. But I will leave that aside for the moment, and comment on the paper as I currently understand it.

          In the third paragraph of the introduction, the authors mention several factors that can lead to "systematic distortion" (which I referred to above as "static distortion"). What is the magnitude of these problems in actual instruments? Were the authors led to this problem because systematic distortion is a problem in their own instrument?

          The authors liken their problem to calibration of the intrinsic camera parameters in computer vision, and refer to "self calibration," with a number of relatively recent references given. I would like to draw the authors' attention to a 1989 paper by Maloney & Ahumada (Maloney, L., & Ahumada, A. J. (1989). Learning by assertion: A method for calibrating a simple visual system. Neural Computation, 1, 387-395) that addresses a related problem, namely how the brain "calibrates" itself to compensate for disarray in the positions of the photoreceptors. The represents an extreme form of systematic distortion, in which each input pixel is displaced in the imaging plane by a random amount. Their solution works by imposing a condition they call "translation invariance," which states that the image of a translated scene should be an equivalent translation of the image of the original scene, or I( T(s,delta) ) = T( I(s), delta ). If I have understood the present paper correctly, the method describe employs the same principle. I think the fact that I am not 100% sure means that the exposition could be clarified somewhat; I suggest starting out with a clear statement of the principle behind the solution (whether it be translation invariance or something else), before delving into the details. The original Maloney & Ahumada work assumed the motion generating the translations was known exactly (perhaps using an efference copy of the eye movement signals, or proprioception); later work explored the performance of the method using imperfectly known motion estimated from the raw uncorrected images themselves, which seems analgous to the problem of calibrating an ocular imaging system.

          I did not find the one-dimensional example particularly helpful. Writing the equations in vector form can make them as concise as a 1-D solution, but then you would be done.

          Here is what I would ideally like to see: apply the method to a more realistic set of synthetic images, generated by taking a large retinal image as ground truth, and then simulating capture through an imaging system containing realistic distortion as might be encountered in practice. First do it for a flood-illuminated system, so that dynamic distortions are not an issue; then simulate dynamic distortions in a scanned system by simulating sampling the image sequentially in time while it is moved according to a realistic eye movement signal (with and without torsion).

          It is hard to know how to interpret the results shown in figure 7; can error bounds be put on the solution? Some statistics might help here.

          -Jeff Mulligan

          Minor points:

          line 50: I would have thought that in the computer vision context, "rotation" of the camera includes pan and tilt, which for small angles translates the image, but is properly a rotation. To use this for camera calibration, it is necessary for the axis of rotation to pass through the optical nodal point, which requires some care.

          line 55: "To our knowledge none of the methods advance have treated the case of dynamic and systematic distortions together..." - this certainly leads one to expect that the authors will treat them together, but that does not seem to be the case.

          line 65: It seems the word "not" may be missing?

          line 79: "... such any one of the ... images " is the word "that" missing after "such"?

          Table 1: The final row appears to be a map from the first "image" to the second, but the legend states the opposite.

          line 190: A little more explanation of the mathematics might be appropriate here. Matrix A is known and is sparse, but has dimensions N x M*N... Apparently the inverse matrix is really a pseudo-inverse? The description of lsqminnorm sounds like it is applied to under-constrained problems... But this problem is not under-constrained?

          line 258: perfect registration is assumed; How robust is the method to registration errors - isn't that important?

          Equation 6: Why is Dx only a function of x, and Dy only a function of y? (compression, but no shear) This appears not to be the case at line 324, where Dx(y) and Dy(y) are referenced?

          Line 459: The polygons are not that easy to compare visually - is it not possible to come up with a quantitative measure?

          Line 498: Why can't the magnitude of the distortion be solved? Is this just saying that there is an unrecoverable constant rigid transformation (like a prismatic shift)?

          Line 543: Instead of referring to data in figure 7 by color, state what the different colored curves represent.

          Line 546: It is assumed that eye movements are negligible compared to the speed of the fast scan in an SLO. This may not be true for saccadic velocities.

          **********

          6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

          If you choose “no”, your identity will remain anonymous but your review may still be made public.

          Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

          Reviewer #1: No

          Reviewer #2: Yes: Jeffrey B. Mulligan

          [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

          While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.


          2 Mar 2021

          Please see attached document "Response to Reviewers.docx".


          20 Apr 2021

          PONE-D-20-39508R1

          Towards distortion-free imaging of the eye

          PLOS ONE

          Dear Dr. Bedggood,

          Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

          While new concerns were raised in this round of review, particularly by reviewer 2, i.e. eye position related distortion, that may help further improve the simulation, you do not have to agree to all the points. A balanced discussion of the potential limitation of the simulation work may help address the review concern.     

          Please submit your revised manuscript by 05/31/2021. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

          Please include the following items when submitting your revised manuscript:

          • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
          • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
          • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

          If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

          If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

          We look forward to receiving your revised manuscript.

          Kind regards,

          Yuhua Zhang

          Academic Editor

          PLOS ONE

          [Note: HTML markup is below. Please do not edit.]

          Reviewers' comments:

          Reviewer's Responses to Questions

          Comments to the Author

          1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

          Reviewer #1: (No Response)

          Reviewer #2: (No Response)

          **********

          2. Is the manuscript technically sound, and do the data support the conclusions?

          The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

          Reviewer #1: Yes

          Reviewer #2: No

          **********

          3. Has the statistical analysis been performed appropriately and rigorously?

          Reviewer #1: N/A

          Reviewer #2: No

          **********

          4. Have the authors made all data underlying the findings in their manuscript fully available?

          The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

          Reviewer #1: Yes

          Reviewer #2: Yes

          **********

          5. Is the manuscript presented in an intelligible fashion and written in standard English?

          PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

          Reviewer #1: Yes

          Reviewer #2: Yes

          **********

          6. Review Comments to the Author

          Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

          Reviewer #1: The authors extensively addressed the reviewers' concerns and the manuscript is well revised.

          I only have a few concerns.

          - Eye motion simulation

          According to the details of the description of the presented simulation, the authors assume that the eye motion (full-field movements + dynamic distortion) has no variation in the horizontal component during single image acquisition (Dx(y) is a constant).

          Even if there is a sinusoidal horizontal dynamic distortion, it is not a realistic simulation, but I think it is better to remind readers of this point.

          - Comparison of 6-cluster outputs with a ground truth

          The authors described the similarity between the outputs and a ground truth that "This can be further verified by the similarity with the ground truth image of Fig 8C (correlation = 0.991; recovering the images drawn from other clusters gave near-identical results) and ..."

          I suggest describing the range of correlation coefficient to support their claim.

          - Fig 8

          "... running on the six outputs of dynamic recovery (one for each cluster), examples of which were shown in B and D."

          Aren't they "D and E"?

          Reviewer #2: In general, the authors have made a good effort to respond to the concerns raised by the original reviews.

          I have one serious concern that did not occur to me when I read the paper for the first time a few months ago: the method explicitly solves for "static" distortion that is the same on every frame. It does this by comparing the positions of features in pairs of retinal images, obtained with the eye in a different pose. The method must assume that either the stationary eye does not contribute any distortion itself, or that whatever distortion it does contribute is constant (does not depend on eye position). I am doubtful that either one of these assumptions hold; in fact, the authors themselves mention distortion contributed by the eye at line 57. This really needs to be addressed. The simulations do not consider distortions that depend on eye position.

          I am trying to imagine what would happen when a distortion-free imaging system is used to acquire retinal imagery from an eye exhibiting position-dependent distortion. I haven't been able to completely wrap my head around it, but I am confident that the result will not be an undistorted image of the retina, because the assumptions of the algorithm are violated. In practice, one approach to this problem could be to try to calibrate the instrument without an eye, by placing a calibration target at the focal plane of the device and moving it. Then when trying to estimate the (possibly position-dependent) distortion added by the eye, the instrument distortion could be factored out.

          Before moving to my comments on the revised manuscript, I would like to respond to a few of the responses to my original review:

          "The problem cannot be vectorized per se, although the distortion maps are represented by column vectors. At any rate, Eq. 2 could not be made any more concise." I certainly agree with the second statement, Eq. 2 is great as it stands, and should be equally true for the two-dimensional problem of interest. One equation that I would like to see in vector form is Eq. 3, which is written out as two equations, one for the x coordinate and another for the y coordinate. The x and y coordinates could be subsumed into a position p, and equation 3 could then be written in a coordinate-free way as:

          f(p_R) - f(p_M) = p_M - R p_M - T

          where R and T are the rotational and translational components of the affine transformation. This form still requires the definition of an origin, i.e. the location of the zero-length position vector that is the center of the rotation. As far as I can tell, this equation should be the same as equation 1. I really think that it the method can be described a little more clearly then the one-dimensional example will not really add anything.

          Further on, it appears that there was a typo in my original review, as evidenced by the authors' response, "This comment is ad odds with the reviewer's comment above, with which we agree, that 'the synthetic distortions that are corrected are arbitrary sinusoidal warps, that to my knowledge are common distortions in any sort of imaging system'." What I meant to say in the original review is that to the best of my knowledge these are NOT common distortions in any sort of imaging system. I apologize for my sloppy proof-reading. That being said, I grant that any arbitrary distortion function can be synthesized from sinusoidal warps via Fourier series, so this doesn't seem like a critical issue; as the sythetic distortions appear to be quite large (possibly larger than what might be encountered in practice), if the algorithm can handle them then it will probably work well in practice. Still, it might be nice to test with a circularly-symmetric distortion function, similar to what might be encountered with a lens.

          In their response to another point, the authors state that, "there is a lack of experimental evidence regarding the degree of distortion expected in real eyes." I would like to note that there is a relationship between geometric distortion and optical aberrations, and there are databases of wave aberration functions measured on collections of eyes, from which it might be possible to make reasonable estimates of the geometric distortions encountered in real eyes. This illustrates a problem with the approach, however, which is perhaps unique to the problem of ocular imaging: instrument-related distortions can safely be assumed to depend only on the sensor pixel location. Eye-related distortions, on the other hand, maybe unique for each different pose of the eye. This seems to me to be a possibly-fatal flaw with the whole approach, given that the method relies upon eye movements to produce the image shifts, and assumes that the distortion function is constant and locked to the sensor. This problem must be addressed, if only by estimating the eye-related distortions from aberration data, and showing that they are insignificant. It also seems possible that they might become significant for highly-aberrated eyes, and this should be considered also.

          I have a serious concern with a statement made on line 473: the authors demonstrate that when only two images are input, which are vertically offset, that the horizontal distortion is not recovered with as much accuracy. The authors conclude that this can be fixed by shifting obliquely. This does not seem intuitively correct to me; it should be possible to formulate the problem in a coordinate-free way, in which case I would expect that larger errors would be encountered in the dimension orthogonal to the shift, regardless of what direction that might happen to be. For an oblique shift, the largest errors should be found in the orthogonal oblique direction. This is just my intuition, of course, but it seems highly unlikely to me that the coordinate axes are special, and this seems even more critical when this principle has been immortalized as part of the name of their procedure.

          Detailed comments:

          Line 65: It is stated that "Distortion is likely to be even higher ... when constructing montages of the retina or imaging with very wide field devices." In this case, it is not clear to me that the problem is distortion, so much as the fact the the retina is not flat, so any attempt to create a montage by translating and registering flat images is doomed to have errors, even in the absence of distortion. The projection of a spherical surface onto a planar image sensor could produce the appearance of distortion when in fact it is an inherent property of the object being imaged, not the imaging system.

          Line 108: "... trial-and-error may be used to modify the system's calibration function ..." - Note that in reference 33 Maloney & Ahumada do not use "trial-and-error," but rather the Widrow-Hoff learning rule, similar to back-propagation as used in neural networks.

          Line 148: The text says that the letter "H" is highlighted in red, but it does not appear so in my review copy. Consider depicting this in a way which will still work when printed in black-and-white.

          Line 159: Xm is referred to as the "moving image", but wouldn't it be more accurate to call it the "translated image" or "moved image"?

          Line 160: Even though it is somewhat clear from context, it would be good to define the symbols used in these equations. It is kind of confusing, because it is stated (line 165) that the expressions that are equated to x_0 "refer to 'world' co-ordinates", but clearly x_R and x_M refer to sensor co-ordinates. It would be better not to use the same symbol (x) to refer to values in different domains.

          Line 168: The symbol D_M,R is said to represent "the output of pixelwise registration between the two images," and is therefore a vector (with dimensionality equal to the number of pixels), while T_M,R is "the full-field translation of the moving image relative to the reference image," which in this example is a scalar or 1-D vector. In the case of 2-dimensional images, if T is allowed to be an arbitrary rigid transformation (e.g. translation + torsional rotation), the T would need to be different for every pixel and would have the same dimensionality as D. This should be clarified. The authors rejected my suggestion to present the equations in vector form, but I still encourage them to adopt type-setting conventions that help indicate the types of various entities, such as bold non-italic lower-case letters for vectors. Perhaps in this case D and T are really functions of x? This needs to be fixed up so that a reader could easily implement the algorithm themselves from the description.

          Line 193: The corresponding pixels are denoted x_1 and x_2, while earlier the were x_R and x_M. These should be consistent.

          Line 217: Why are these equations relegated to a figure instead of typeset in-line, as was done for the previous equations?

          Figure 2: There is something that I don't understand about this: x_1, x_2, etc, which label the columns of A, refer to locations on the sensor, while the vector f is the unknown distortion map that we wish to obtain. Why should there only be rows relating successive locations, such as x_1 and x_2, x_2 and x_3, etc? It seems like x_1 could correspond to any of the other x's for an appropriate motion. This doesn't make a lot of sense to me.

          Equation 3 (line 292): T represents an affine transformation matrix, but it is written as a function with two arguments. In equation 4, there appear to be 3 parameters (theta, T_x, and T_y). What are the two arguments in equation 3?

          Line 330: "Ground truth" images were obtained from a flood-illumination ophthalmoscope, and subsequently distorted with known distortions. While mostly irrelevant for the work presented, I am wondering whether the authors applied their method to solve for the (hopefully small) distortions in their ophthalmoscope?

          Line 439: It is not at all clear to me that the static and dynamic distortions can be treated independently; it seems likely to me that the presence of static distortions will introduce small errors in the estimation of the dynamic distortions, although these will probably be quite small for small movements, and slowly-varying static distortions. This could be explored in simulation.

          Line 600: "The method is blind to the absolute value of the distorting function..." - I think that "absolute value" is not the best term to use here. A constant translational shift is given here as an example, but isn't that in fact the only component of distortion that falls into this category? I would rephrase it to say that the method is blind to prismatic shifts, or translations that are constant over the whole image. On second thought, I think that this also includes uniform magnification/minification; if this is true, then why not mention it as well?

          Lines 605-611: I'm not sure what is meant by "'islands' of connected pixels". I think what might be meant here is that however many images are collected, there need to be some that place a register-able feature at each image location. Obviously if the target (retina) has featureless regions, then there will be nothing to register and therefore no data to input to the algorithm. But with normal eyes, I don't see how this would ever arise in practice?

          line 618: Another factor that could produce intensity differences between frames is vignetting produced in certain poses of the eye.

          lines 622-634: The paragraph discusses a two-step procedure for dealing with simultaneous static and dynamic distortions. It is not clear to me that static distortions won't introduce biases into the solution for the dynamic distortions, and that these biases won't in turn produce error in the DIOS process. The way that I conceptualize the problem of dynamic distortion is that there is a true underlying undistorted image, with N_rows x N_cols scalar parameters (larger than a single acquired image, as this represents the correctly registered mosaic of all of the images), and three pose parameters for each frame, parameterizing the eye pose (rotation). Each image contributes M_rows x M_cols data points (the pixel intensities in a single acquired frame), and 3 additional unknowns (the pose parameters). So it is clear that as we add images, the number of inputs grows much faster than the number of unknowns. But I don't know of a practical way to perform the optimization to obtain the solution (although there certainly may be one). Existing methods might provide a good initial solution that could be further refined. Incorporating static distortion into this model adds an additional 2 x M_rows x M_cols parameters (the distortion translation vector for each pixel), but as this is the same for every frame it does not increase as frames are added, it just makes the optimization a little more complicated. Even if it is too slow to be useful in practice, if such a method could be used to generate a solution, I would have more confidence in it than sequential optimization of sub-problems, and it would be.

          The problem discussed in the preceding paragraph is mentioned by the authors themselves at line 643, where they say, "It is noteworthy that although [sic] the residual static correction obtained was not simply equal to the underlying static correction applied." I'm not sure what this means, is the "underlying static correction applied" the same as the synthetic distortion applied, i.e. ground truth?

          Lines 688-693: The authors discuss various ways to make the computation of the solution more efficient. One approach that they may wish to consider is a progressive "coarse-to-fine" solution, where the distortion map is built up from progressively higher frequency components, kind of like solving for the Fourier components of the distortion map sequentially. The advantage is that the lower frequency components might be solved for using subsampled data (i.e., a "pyramid" approach), reducing the amount of computation required. This suggestion is close to the approach suggested at line 690, where they suggest down-sampling by "binning."

          **********

          7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

          If you choose “no”, your identity will remain anonymous but your review may still be made public.

          Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

          Reviewer #1: No

          Reviewer #2: Yes: Jeffrey B. Mulligan

          [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

          While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.


          9 May 2021

          Please see attached "Response to Reviewers" document.


          25 May 2021

          Towards distortion-free imaging of the eye

          PONE-D-20-39508R2

          Dear Dr. Bedggod, 

          We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

          Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

          An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

          If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

          Kind regards,

          Yuhua Zhang

          Academic Editor

          PLOS ONE

          Additional Editor Comments (optional):

          Reviewers' comments:


          2 Jun 2021

          PONE-D-20-39508R2

          Towards distortion-free imaging of the eye

          Dear Dr. Bedggood:

          I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

          If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

          If we can help with anything else, please email us at plosone@plos.org.

          Thank you for submitting your work to PLOS ONE and supporting open access.

          Kind regards,

          PLOS ONE Editorial Office Staff

          on behalf of

          Dr. Yuhua Zhang

          Academic Editor

          PLOS ONE

          https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
          https://www.researchpad.co/tools/openurl?pubtype=article&doi=10.1371/journal.pone.0252876&title=Towards distortion-free imaging of the eye&author=Phillip Bedggood,Andrew Metha,Yuhua Zhang,&keyword=&subject=Research Article,Biology and Life Sciences,Anatomy,Head,Eyes,Medicine and Health Sciences,Anatomy,Head,Eyes,Biology and Life Sciences,Anatomy,Ocular System,Eyes,Medicine and Health Sciences,Anatomy,Ocular System,Eyes,Biology and Life Sciences,Physiology,Sensory Physiology,Visual System,Eye Movements,Biology and Life Sciences,Neuroscience,Sensory Systems,Visual System,Eye Movements,Research and Analysis Methods,Imaging Techniques,Biology and Life Sciences,Cell Biology,Cellular Types,Animal Cells,Neurons,Afferent Neurons,Photoreceptors,Biology and Life Sciences,Neuroscience,Cellular Neuroscience,Neurons,Afferent Neurons,Photoreceptors,Biology and Life Sciences,Neuroscience,Cognitive Science,Cognitive Psychology,Perception,Sensory Perception,Sensory Receptors,Photoreceptors,Biology and Life Sciences,Psychology,Cognitive Psychology,Perception,Sensory Perception,Sensory Receptors,Photoreceptors,Social Sciences,Psychology,Cognitive Psychology,Perception,Sensory Perception,Sensory Receptors,Photoreceptors,Biology and Life Sciences,Neuroscience,Sensory Perception,Sensory Receptors,Photoreceptors,Biology and Life Sciences,Cell Biology,Signal Transduction,Sensory Receptors,Photoreceptors,Physical Sciences,Mathematics,Applied Mathematics,Algorithms,Research and Analysis Methods,Simulation and Modeling,Algorithms,Engineering and Technology,Equipment,Optical Equipment,Adaptive Optics,Biology and Life Sciences,Neuroscience,Cognitive Science,Cognitive Psychology,Perception,Sensory Perception,Vision,Biology and Life Sciences,Psychology,Cognitive Psychology,Perception,Sensory Perception,Vision,Social Sciences,Psychology,Cognitive Psychology,Perception,Sensory Perception,Vision,Biology and Life Sciences,Neuroscience,Sensory Perception,Vision,Biology and Life Sciences,Anatomy,Ocular System,Ocular Anatomy,Retina,Medicine and Health Sciences,Anatomy,Ocular System,Ocular Anatomy,Retina,